INTRODUCTION

UNPACKING
Packing List and verify that you have received all equipment. If there are any questions about the shipment, use the phone numbers listed on the back cover to contact the Customer Service Department nearest you.

Upon receipt of shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.

Note: The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.

Verify that you receive the following items in the shipping box:

QTY
DESCRIPTION
1
Programmable indicator/controller with all applicable connectors attached.
1
Owner's Manual
1
Quick Start Manual

Note:
If you ordered any of the available options (except the ôBLö blank Lens option), they will be shipped in a separate container to avoid any damage to your indicator/controler.

Caution SAFETY CONSIDERATIONS
This device is marked with the international Caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).

Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the unit if damage is found.

This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power connections). Installation of this instrument should be done by Qualified personnel. In order to ensure safe operation, the following instructions should be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947û1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.

Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
  • Caution The Protective Conductor must be connected for safety reasons. Check that the power cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected.

  • Do not exceed voltage rating on the label located on the top of the instrument housing.

  • Always disconnect power before changing signal and power connections.

  • Do not use this instrument on a work bench without its case for safety reasons.

  • Do not operate this instrument in flammable or explosive atmospheres.

  • Do not expose this instrument to rain or moisture.
EMC Considerations
  • Whenever EMC is an issue, always use shielded cables.

  • Never run signal and power wires in the same conduit.

  • Use signal wire connections with twisted-pair cables.

  • Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.

 Home Home  Table of Contents TOC  Back Back  Next Next Help Price List Specifications Print Manual Manuals Home